CVE-2015-1012
No known exploitation. EPSS puts it in the 54th percentile. Only a workaround so far.
What to do
No fixed build is published yet. The vendor describes a workaround.
What it is
From the CVE record
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
In plain language
No plain-language summary for this CVE yet.
Exploitation
Where each signal puts this CVE on the scale from published to confirmed exploited.
- CISA KEV
Not in the catalog. CISA has not confirmed exploitation.
- Public exploits
No public exploit or proof of concept found in the sources we track.
- EPSS
0.8% chance of exploitation activity in the next 30 days, which ranks it in the 54th percentile of scored CVEs.
Exploit Prediction Scoring System, FIRST.org. A probability, not a confirmation.
Lifecycle
5 events over 2068 days, from the signal feeds we watch.
- Workaround availablerecord updated
- Publishedweakness classified, att&ck mapped
Affected products
Technical detail
CVSS 3.0 vector
Open in the CVSS calculatorCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Scored 7.5 by NVD.
How it is reached
- Attack Vector NetworkExploitable remotely over the network without any special conditions
- Attack Complexity LowNo special conditions — the attack can be reliably reproduced
- Privileges Required NoneNo authentication required — anyone can exploit this
- User Interaction NoneNo user interaction needed — fully automated exploitation
Scope
- Scope UnchangedImpact is limited to the vulnerable component itself
Impact if exploited
- Confidentiality HighTotal information disclosure — all data in the component is compromised
- Integrity NoneNo integrity impact
- Availability NoneNo availability impact
Weaknesses
ATT&CK techniques
Mapped from the weaknesses above (CWE to ATT&CK), not observed in attacks.
- CollectionT1005Data from Local Systemmedium confidence
- Credential AccessT1552Unsecured Credentialshigh confidence
Sources
Watch the software you run.
My Stack ranks new CVEs for your products by real-world exploitation, so the next serious one reaches you without reading every advisory.
We'll flag the next CVE, public exploit or patch for Hospira, not every advisory.
A free account adds
- The full version matrix and every affected product
- Exploit links, proofs of concept and Metasploit modules
- Email alerts for the products you watch
- The same data over REST API, MCP and CLI